ISO 11140-4:2007

Sterilization of health care products -- Chemical indicators -- Part 4: Class 2 indicators as an alternative to the Bowie and Dick-type test for detection of steam penetration

Üldinfo

Kehtiv alates 15.03.2007
Direktiivid või määrused
puuduvad

Standardi ajalugu

Staatus
Kuupäev
Tüüp
Nimetus
15.03.2007
Põhitekst
Põhitekst
ISO 11140-4:2007 specifies the performance for a Class 2 indicator to be used as an alternative to the Bowie and Dick-type test for steam sterilizers for wrapped health care goods (instruments, etc. and porous loads).
An indicator complying with ISO 11140-4:2007 incorporates a specified material which is used as a test load. This test load may, or may not, be re-usable. ISO 11140-4:2007 does not specify requirements for the test load, but specifies the performance of the indicator in combination with the test load with which it is intended to be used. The indicator specified in ISO 11140-4:2007 is intended to identify poor steam penetration but does not necessarily indicate the cause of this poor steam penetration.
This part of ISO 11140-4:2007 does not include test methods to establish the suitability of these indicators for use in sterilizers in which the air removal stage does not include evacuation below atmospheric pressure.