EVS-EN IEC 61010-2-032:2021+A11:2021

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-032: Particular requirements for hand-held and hand-manipulated current sensors for electrical test and measurement (IEC 61010-2-032:2019 + COR1:2020)

General information

Valid from 15.02.2022
Base Documents
IEC 61010-2-032:2019; IEC 61010-2-032:2019/COR1:2020; EN IEC 61010-2-032:2021; EN IEC 61010-2-032:2021/A11:2021
Directives or regulations
2014/35/EU Low voltage (LVD)

Standard history

Status
Date
Type
Name
Main
prEN IEC 61010-2-032:2022
15.02.2022
Main + amendment
01.12.2021
Main
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replace the existing text with the following:
This part of IEC 61010 specifies safety requirements for HAND-HELD and hand-manipulated current sensors described below.
These current sensors are for measuring, detecting or injecting current, or indicating current waveforms on circuits without physically opening the current path of the circuit being measured.
They can be stand-alone current sensors or accessories to other equipment or parts of combined equipment (see Figure 101). These include measurement circuits which are part of electrical test and measurement equipment, laboratory equipment, or process control
equipment. These current sensors and circuits need additional protective means between the current sensor, the circuit and an OPERATOR.
NOTE 1 Combined equipment is equipment that is electrically connected to a current sensor by means of a permanent connection which can be detached only by the use of a TOOL.
NOTE 2 Some current sensors are also known as current clamps, CLAMP MULTIMETERS and current probes.
Current sensors are hand-manipulated before and/or after a test or measurement, but do not
necessarily need to be HAND-HELD during the test or measurement. Current sensors used as FIXED EQUIPMENT are not within the scope of this document.
The following types of current sensors are covered:
a) Type A: a current sensor designed to be applied to or removed from HAZARDOUS LIVE UNINSULATED CONDUCTORS. Type A current sensors have defined HAND-HELD or handmanipulated parts providing protection against electric shock from the conductor being measured, and also have protection against short -circuits between wires and between
busbars during clamping.
b) Type B: a current sensor which has protection against short -circuits between wires or busbars during clamping but without defined HAND-HELD or hand-manipulated parts which provide protection against electric shock during clamping. Additional protective means are
necessary to avoid electric shock from HAZARDOUS LIVE conductors which cannot be deenergised during application or removal of the current sensor.
EXAMPLE 1 Flexible current sensors.
c) Type C: a current sensor without protection against short -circuits between wires or busbars during clamping. Type C current sensors are intended to be applied to or removed from HAZARDOUS LIVE UNINSULATED CONDUCTORS or from non-limited-energy circuit conductors only when they are de-energised.
EXAMPLE 2 Split-core transducers.
d) Type D: a current sensor designed to be applied to or removed from insulated conductors or from limited-energy circuit conductors.
EXAMPLE 3 Current probes for oscilloscopes and earth leakage current detectors.
All current sensors can also be used with insulated conductors. In this case, HAZARDS are limited to acceptable levels by the insulation of the conductors.
Additional requirements for CLAMP MULTIMETERS are given in Annex EE.
Figure 101 shows graphical representations of typical current sensors for illustration purposes.
Current sensors can look different depending on the design.
1.2.1 Aspects included in scope
Add the following three new paragraphs at the end of the subclause:
Requirements for protection against HAZARDS resulting from NORMAL USE and REASONABLY
FORESEEABLE MISUSE of measuring circuits are given in Clause 101.
Requirements for prevention of HAZARD from arc flash and short-circuits are given in Clause 102.
Requirements for reliance on the displayed value of CLAMP MULTIMETERS are given in Clause EE.5 .

Required fields are indicated with *

*
*
*
PDF
31.72 € incl tax
Paper
31.72 € incl tax
Browse standard from 2.44 € incl tax
Standard monitoring

Customers who bought this item also bought

Main + amendment

EVS-EN IEC 61010-2-030:2021+A11:2021

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-030: Particular requirements for equipment having testing or measuring circuits (IEC 61010-2-030:2017)
Newest version Valid from 03.05.2021
Amendment

EVS-EN 61010-1:2010/A1:2019

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
Newest version Valid from 01.03.2019
Main

EVS-EN 61010-1:2010

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements
Newest version Valid from 06.12.2010
Main

EVS-EN 61869-1:2009

Instrument transformers -- Part 1: General requirements
Newest version Valid from 06.11.2009