201.1 Scope,and object
Clause 1 of the general standard1 applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS, hereafter referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This particular standard does not apply to
– headlights;
– endoscopes, laparoscopes and their light sources, which are covered by IEC 60601-2-18;
– luminaires used in dentistry, which are covered by ISO 9680;
– luminaires for general purposes, which are covered by IEC 60598-2-1 and IEC 60598-2-4;
– luminaires dedicated to therapeutic purposes;
– special purpose lights with different conditions of use such as light sources intended solely for decontamination of air and surfaces, UV lights for dermatological diagnosis, slit lamps for ophthalmology, lights for surgical microscopes and lights for surgical navigation systems;
– lights connected to surgical instruments, such as luminous retractors;
– luminaires for emergency lighting, which are covered by IEC 60598-2-22.
NOTE See also 4.2 of the general standard.
SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS are medical devices and not general lighting equipment.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE requirements for SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS as defined in 201.3.