Skip to main content
Back

EVS-EN 14626:2012

Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy

General information

Valid from 03.10.2012
Base Documents
EN 14626:2012
Directives or regulations
None

Standard history

Status
Date
Type
Name
Main
prEN 14626
03.10.2012
Main
This European Standard specifies a continuous measurement method for the determination of the concentration of
carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle.
This standard describes the performance characteristics and sets the relevant minimum criteria required to select
an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the
evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements
as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality
assurance for use.
The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air
up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type
approval test.
NOTE 1 Other ranges may be used depending on the levels present in ambient air.
NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges
and uncertainty requirements may not apply.
The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural
areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources.
The results are expressed in mg/m3 (at 20 °C and 101,3 kPa).
NOTE 3 100 mg/m3 of CO corresponds to 86 µmol/mol of CO.
This standard contains information for different groups of users.
Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide
measurement by non-dispersive infrared spectroscopic analyser and sampling equipment.
Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval
testing of carbon monoxide analysers. These sections contain information about:
 type-approval test conditions, test procedures and test requirements;
 analyser performance requirements;
 evaluation of the type-approval test results;
 evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the typeapproval
test results.
Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of
carbon monoxide in ambient air. These sections contain information about:
 initial installation of the analyser in the monitoring network and acceptance testing;
 ongoing quality assurance/quality control;
 calculation and reporting of measurement results;
 evaluation of the uncertainty of measurement results under practical monitoring conditions.

Required fields are indicated with *

*
*
*
PDF
39.04 € incl tax
Paper
39.04 € incl tax
Browse standard from 2.44 € incl tax
Standard monitoring

Customers who bought this item also bought

Main

EVS-EN 14211:2012

Ambient air - Standard method for the measurement of the concentration of nitrogen dioxide and nitrogen monoxide by chemiluminescence
Newest version Valid from 03.10.2012
Main + corrigendum

EVS-EN 14212:2012

Ambient air - Standard method for the measurement of the concentration of sulphur dioxide by ultraviolet fluorescence
Newest version Valid from 03.10.2012
Main

EVS-EN 14625:2012

Ambient air - Standard method for the measurement of the concentration of ozone by ultraviolet photometry
Newest version Valid from 03.10.2012
Main

EVS-EN 14662-3:2015

Ambient air - Standard method for the measurement of benzene concentrations - Part 3: Automated pumped sampling with in situ gas chromatography
Newest version Valid from 05.01.2016