Skip to main content

Information service

11 Health care technology
New standards
EVS-EN ISO 18113-1:2024
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 1: Terms, definitions, and general requirements (ISO 18113-1:2022)
Scope: This document defines concepts, establishes general principles, and specifies essential requirements for information supplied by the manufacturer of IVD medical devices.
This document does not address language requirements since that is the domain of national laws and regulations.
This document does not apply to:
a) IVD medical devices for performance evaluation (e.g. for investigational use only);
b) shipping documents;
c) material safety data sheets / Safety Data Sheets;
d) marketing information (consistent with applicable legal requirements).
.
Base documents: ISO 18113-1:2022; EN ISO 18113-1:2024
EVS-EN ISO 18113-4:2024
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 4: In vitro diagnostic reagents for self-testing (ISO 18113-4:2022)
Scope:

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators, and controls intended for self-testing.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for professional use.

Base documents: ISO 18113-4:2022; EN ISO 18113-4:2024
EVS-EN ISO 18113-5:2024
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 5: In vitro diagnostic instruments for self-testing (ISO 18113-5:2022)
Scope: This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for self-testing.
This document is also applicable to apparatus and equipment intended to be used with IVD instruments for self-testing.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for professional use.
Base documents: ISO 18113-5:2022; EN ISO 18113-5:2024
EVS-EN 556-1:2024
Sterilization of medical devices - Requirements for medical devices to be designated "STERILE" - Part 1: Requirements for terminally sterilized medical devices
Scope: This document specifies the requirements for a terminally sterilized medical device to be designated ‘STERILE’. Part 2 of this European standard specifies the requirements for an aseptically processed medical device to be designated "STERILE".
NOTE  For the purpose of the EU Directive(s) for medical devices (see Bibliography), designation of a medical device as ‘STERILE’ is only permissible when a validated sterilization process has been applied. Requirements for validation and routine control of processes for the sterilization of medical devices are specified in EN ISO 11135, EN ISO 11137, EN ISO 14160, EN ISO 14937, EN ISO 17665-1, EN ISO 20857, EN ISO 25424 and ISO 22441.
Base documents: EN 556-1:2024
Replaces: EVS-EN 556-1:2002
ISO 6559:2024
Traditional Chinese medicine — Sterile three-edge needle for single use
Scope: This document specifies the configuration, dimensions, parameters, requirements and test methods of a sterile three-edge needle for single use. It also specifies requirements for packaging, labelling, transit and storage.
This document excludes requirements for the handle of a three-edge needle, which is structurally different from the body of a three-edge needle.
Base documents:
ISO 5910:2024
Cardiovascular implants and extracorporeal systems — Cardiac valve repair devices
Scope: 1.1 This document specifies an approach for verifying and validating the design and manufacture of a heart valve repair system through risk management. The selection of appropriate verification and validation tests and methods are derived from the risk assessment. The tests include assessments of the physical, chemical, biological and mechanical properties of components and materials of heart valve repair systems. The tests also include preclinical in vivo evaluation and clinical investigation of the finished heart valve repair system to assess the safety and effectiveness of the heart valve repair system.
NOTE            For the purposes of this document, effectiveness end point includes clinical performance and benefits.
1.2 This document defines operational conditions and performance requirements for heart valve repair systems where adequate scientific and/or clinical evidence exists for their justification. It also describes the labels and packaging of the device.
1.3 This document applies to all heart valve repair systems that have an intended use to repair and/or improve the function of native human heart valves by acting either on the valve apparatus or on the adjacent anatomy (e.g. ventricle, coronary sinus).
1.4 This document does not apply to cardiac resynchronization therapy (CRT) devices, paravalvular leakage closure devices, systems that do not leave an implant in place (e.g. ablation, radio frequency annuloplasty), apical conduits and devices with components containing viable cells. This document also excludes materials not intended for repairing and/or improving the function of human heart valves according to its intended use (e.g. patch material and sutures used in general surgical practice).
NOTE            A rationale for the provisions of this document is given in Annex A.
Base documents:
Replaces: ISO 5910:2018
ISO 15098:2024
Dentistry — Dental tweezers
Scope: This document specifies general requirements and test methods for metallic dental tweezers of Meriam type and College type.
This document is not applicable to anatomical tweezers and surgical tweezers.
Base documents:
Replaces: ISO 15098:2020
ISO 80369-2:2024
Small-bore connectors for liquids and gases in healthcare applications — Part 2: Connectors for respiratory applications
Scope: This document specifies the design and dimensions for two small-bore connectors intended to be used for connections in respiratory applications of medical devices and accessories. One connector (R1) is intended for use on medical devices and accessories subjected to pressures up to 15 kPa (e.g. a breathing system). The other connector (R2) is intended for use on medical devices and accessories subjected to higher pressures between 15 kPa and 600 kPa (e.g. oxygen therapy tubing).
NOTE 1        The pressure is related to pressure available at the source to which the medical device is connected.
NOTE 2        The intended application does not preclude the use of other connectors on medical devices or accessories within this application.
NOTE 3        Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in device-specific standards.
NOTE 4        If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.
Base documents:
IEC 60601-2-37:2024
Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Scope: IEC 60601-2-37:2024 is available as IEC 60601-2-37:2024 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-37:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.7.2.13. This document does not cover ultrasonic therapeutic equipment. Equipment used for the imaging or diagnosis of body structures by ultrasound in conjunction with other medical procedures is covered. IEC 60601-2-37:2024 cancels and replaces the second edition published in 2007 and Amendment 1:2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) technical and editorial changes resulting from the amended general standard IEC 60601 1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and its collateral standards IEC 60601-1-xx,
b) technical and editorial changes as a result of maintenance to normative references;
c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics standards. In particular, Clause 201.11 about protection against excessive temperatures and other hazards has been fully revised.
Base documents:
Replaced standards
EVS-EN ISO 18113-4:2011
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 4: In vitro diagnostic reagents for self-testing (ISO 18113-4:2009)
Scope: This part of ISO 18113 specifies requirements for information supplied by the manufacturer of IVD reagents for self-testing. This part of ISO 18113 also applies to information supplied by the manufacturer with calibrators and control materials intended for use with IVD medical devices for self-testing. This part of ISO 18113 can also be applied to accessories. This part of ISO 18113 applies to the labels for outer and immediate containers and to the instructions for use. This part of ISO 18113 does not apply to: a) IVD instruments or equipment, b) IVD reagents for professional use.
Base documents: ISO 18113-4:2009; EN ISO 18113-4:2011
EVS-EN ISO 18113-5:2011
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 5: In vitro diagnostic instruments for self-testing (ISO 18113-5:2009)
Scope: This part of ISO 18113 specifies requirements for information supplied by the manufacturer of IVD instruments for self-testing. This part of ISO 18113 also applies to apparatus and equipment intended to be used with IVD instruments for self-testing. This part of ISO 18113 can also be applied to accessories. This part of ISO 18113 does not apply to a) instructions for instrument servicing or repair, b) IVD reagents, including calibrators and control materials for use in control of the reagent, c) IVD instruments for professional use.
Base documents: ISO 18113-5:2009; EN ISO 18113-5:2011
EVS-EN ISO 18113-1:2011
In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 1: Terms, definitions and general requirements (ISO 18113-1:2009)
Scope: This part of ISO 18113 defines concepts, establishes general principles and specifies essential requirements for information supplied by the manufacturer of IVD medical devices. This part of ISO 18113 does not address language requirements, since that is the domain of national laws and regulations. This part of ISO 18113 does not apply to a) IVD devices for performance evaluation (e.g., for investigational use only), b) instrument marking, c) material safety data sheets.
Base documents: ISO 18113-1:2009; EN ISO 18113-1:2011
IEC 60601-2-37:2007/AMD1:2015
Amendment 1 - Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Scope:
Base documents:
IEC 60601-2-37:2007+AMD1:2015 CSV
Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Scope: IEC 60601-2-37:2007+A1:2015 Applies to the basic safety and essential performance of ultrasonic diagnostic equipment. The approach and philosophy used in drafting this particular standard for safety of ultrasonic diagnostic equipment are consistent with those in standards of the IEC 60601 2 xx series that apply to other diagnostic modalities, such as X-ray equipment and magnetic resonance systems. In each case, the safety standard is intended to require increasing sophistication of output display indicators and/or controls with increasing energy levels in the interrogating field of diagnosis. Thus, for all such diagnostic modalities, it is the responsibility of the operator to understand the risk of the output of the ultrasonic diagnostic equipment, and to act appropriately in order to obtain the needed diagnostic information with the minimum risk to the patient. This edition combines the previous edition and its amendments into a form compatible with the parent IEC 60601-1:2005.                        

This consolidated version consists of the second edition (2007)  
and its amendment 1 (2015). Therefore, no need to order amendment in
addition to this publication.
Base documents:
ISO 5910:2018
Cardiovascular implants and extracorporeal systems -- Cardiac valve repair devices
Scope: 1.1    This document applies to all heart valve repair systems that have an intended use to repair and/or improve the function of native human heart valves by acting either on the valve apparatus or on the adjacent anatomy (e.g. ventricle, coronary sinus).
1.2    This document outlines an approach for verifying/validating the design and manufacture of a heart valve repair system through risk management. The selection of appropriate verification/validation tests and methods are derived from the risk assessment. The tests include assessments of the physical, chemical, biological, and mechanical properties of components and materials of heart valve repair systems. The tests also include preclinical in vivo evaluation and clinical investigation of the finished heart valve repair system to assess the safety and effectiveness of the heart valve repair system.
NOTE       For the purposes of this document, effectiveness endpoint includes clinical performance and benefits.
1.3    This document defines operational conditions and performance requirements for heart valve repair systems where adequate scientific and/or clinical evidence exists for their justification.
1.4    This document excludes Cardiac Resynchronization Therapy (CRT) devices, paravalvular leakage closure devices, systems that do not leave an implant in place (e.g. ablation, radio frequency annuloplasty), apical conduits and devices with components containing viable cells. This Standard also excludes materials not intended for repairing and/or improving the function of human heart valves according to its intended use (e.g. patch material and sutures used in general surgical practice).
NOTE       A rationale for the provisions of this document is given in Annex A.
Base documents:
Replaced: ISO 5910:2024
IEC 60601-2-37:2007
Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Scope: Applies to the basic safety and essential performance of ultrasonic diagnostic equipment.

The approach and philosophy used in drafting this particular standard for safety of ultrasonic diagnostic equipment are consistent with those in standards of the IEC 60601 2 xx series that apply to other diagnostic modalities, such as X-ray equipment and magnetic resonance systems.

In each case, the safety standard is intended to require increasing sophistication of output display indicators and/or controls with increasing energy levels in the interrogating field of diagnosis. Thus, for all such diagnostic modalities, it is the responsibility of the operator to understand the risk of the output of the ultrasonic diagnostic equipment, and to act appropriately in order to obtain the needed diagnostic information with the minimum risk to the patient.

This second edition cancels and replaces the first edition published in 2001 and its Amendment 1 (2004) and Amendment 2 (2005). This edition combines the previous edition and its amendments into a form compatible with the parent IEC 60601-1:2005.
Base documents:
EVS-EN 60522:2002
Determination of the permanent filtration of X-ray tube assemblies
Scope: This standard defines the concept of permanent filtration in X-ray tube assemblies for medical diagnosis and radiotherapy and describes a method for its determination. It contains requirements for statements of compliance for accompanying documents and for marking on X-ray tube assemblies.
Methods are given to determine the permanent filtration in an X-ray tube assembly with an accuracy that is sufficient to enable the appropriate additional filtration to be provided in order to attain the desired total filtration.
Base documents: IEC 60522:1999; EN 60522:1999
EVS-EN 556-1:2002
Sterilization of medical devices - Requirements for medical devices to be designated "STERILE" - Part 1: Requirements for terminally sterilized medical devices.
Scope: This European Standard specifies the requirements for a terminally-sterilized medical device to be designated 'STERILE': Part 2 of this European Standard specifies the requirements for an aseptically processed medical device to be designated "STERILE".
Base documents: EN 556-1:2001; EN 556-1:2001/AC:2006
Replaced: EVS-EN 556-1:2024
ISO 15098:2020
Dentistry -- Dental tweezers
Scope: This document specifies general requirements and test methods for metallic dental tweezers of the Meriam type and for College type.
This document is not applicable to anatomical tweezers and surgical tweezers.
Base documents:
Replaced: ISO 15098:2024
Drafts
prEN 16128
Ophthalmic optics - Reference method for the testing of spectacle frames and sunglasses for nickel release
Scope: Specification of the reference method for the testing of spectacle frames, ready-to-wear near-vision spectacles, sunglasses and spectacle frames used for eye and face protection for nickel release.
The reference method supports the demonstration of conformity with the limit value for nickel release of 0,5 μg·cm-2·week-1 set forth by European Regulation [Regulation (EC) No 1907/2006, REACH, in particular Commission Regulation (EC) No 552/2009 of 22 June 2009 amending regulation (EC) No 1907/2006 of the European Parliament and of the Council on the Registration, Evaluation,
Authorization and restriction of Chemicals (REACH) as regards Annex XVII RESTRICTIONS ON THE MANUFACTURE, PLACING ON THE MARKET AND USE OF CERTAIN DANGEROUS SUBSTANCES, PREPARATIONS AND ARTICLES].
The standard applies to those parts of metal spectacle frames and those metal parts of combination spectacle frames that are intended to come into direct and prolonged contact with the skin of the wearer. It also applies to those relevant metal parts of ready-towear near-vision spectacles, sunglasses and spectacle frames used for eye and face protection.
Base documents: prEN 16128
prEN ISO 10993-1
Biological evaluation of medical devices - Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process (ISO/DIS 10993-1:2024)
Scope: This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk management process;
—  the general categorization of medical devices based on the nature and duration of their contact with the body;
—  the evaluation of existing relevant data from all sources;
—  the identification of gaps in the available data set on the basis of a risk analysis;
—  the identification of additional data sets necessary to analyse the biological safety of the medical device;
—  the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct or indirect contact with:
—  the patient's body during intended use;
—  the user's body, if the medical device is intended for protection (e.g., surgical gloves, masks and others).
This document is applicable to biological evaluation of all types of medical devices including active, non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
—  risks, such as changes to the medical device over time, as a part of the overall biological safety assessment;
—  breakage of a medical device or medical device component which exposes body tissue to new or novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Device-specific or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents and other pathogens.
Base documents: ISO/DIS 10993-1; prEN ISO 10993-1
prEN ISO 18374
Dentistry - Artificial intelligence (AI) and augmented intelligence (AuI) based 2D radiograph analysis - Data generation, data annotation and data processing (ISO/DIS 18374:2024)
Scope: This document applies to software for 2D x ray image analysis in dental medicine, the basis of which is the application of artificial intelligence (AI).
The scope of applicability of this document includes
a) Software-as-a-Medical Device (SaMD)
b) Software-in-a-Medical Device (SiMD) and
c) Software to improve the efficiency of healthcare applications.
This document applies to AI applications that learn statically and continuously/incrementally.
Base documents: ISO/DIS 18374; prEN ISO 18374
prEN ISO 11980
Ophthalmic optics - Contact lenses and contact lens care products - Guidance for clinical investigations (ISO/DIS 11980:2024)
Scope: ISO 11980:2012 gives guidelines for the clinical investigation of the safety and performance of contact lenses and contact lens care products.
Base documents: ISO/DIS 11980; prEN ISO 11980
prEN IEC 80601-2-31:2024
Medical electrical equipment - Part 2-31: Particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source
Scope: Clause 1 of the general standard1 applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of EXTERNAL
PACEMAKERS powered by an INTERNAL ELECTRICAL POWER SOURCE, hereafter referred to as ME EQUIPMENT.
This document applies to PATIENT CABLES as defined in 201.3.209, but does not apply to LEADS as defined in 201.3.206.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document does not apply to the implantable parts of ACTIVE IMPLANTABLE MEDICAL DEVICES covered by ISO 14708-1. This document does not apply to EXTERNAL PACEMAKERS which can be connected directly or indirectly to a SUPPLY MAINS.
This document does not apply to transthoracic and oesophageal pacing ME EQUIPMENT and antitachycardia ME EQUIPMENT.
Base documents: 62D/2136/CDV; prEN IEC 80601-2-31:2024